Description
- Clotting : Optical clot detection with 6 LEDs
- Chromogenic: Absorbance with 1 LED
- Immunology: Turbidimetry with 1 LED
- Up to 140 PT/hour
- Combined assays around 50 to 60 tests/hour
- Samples: Plasma, serum
- Routine and STAT processing
Reagents / samples carousel
- Sample carousel : 40 sample positions for primary tubes 13x75mm (7, 5, 3.5, 3, 2, 1 ml), microtubes
- Reagent carousel: Specific coagulation carousel for bottles and tubes (1.5ml, 5ml and 10ml) 35 positions
- Carousel refrigeration
- Sample dead volume: 50 µl
- Positive identification for samples and reagents by internal barcode reader
Pipetting system
- 2 precision maintenance free pumps with ceramic pistons
- Reagent and sample dispensing: 2-220 µl with 0.1µl increments
- Mixing system on the sample probe
- Preheating system on the reagent probe
Reactions
- Reaction support: 96 well microplate
- Microplate heating system: 37°C
- Reaction volume: 105 – 255 µl
Optical system
- Six 405nm LEDs for chronometric determinations
- One 656nm LED with filter for colorimetric determinations
- One 405nm LED with filter for colorimetric determinations
- Resolution: 14 bits
Other
- Power: 95-230 Volts, 50-60 Hz
- Dimensions: 850 x 660 x 550 mm
- Weight: 60 kg
- CE Certification
Current MAXMAT test offer
Coming soon: (open for other parameters on the market):
- Thrombin Time
- D-Dimer
- Factors
- Chromogenic: AT, PC, PS, Heparin and LMWH, antiplasmin, plasminogen
- Latex: PC, total and free PS, vWF antigen
- APCr, PTT LA